Laser Test Stand Operation

==============================================================================

TEST/PROTO.04 JAN 12, 2000

'PROTO' - Laser scan software for silicon microstrip detectors

==============================================================================

0.0 Table of Contents

0. Table of Contents

1. Test Overview

2. Safety Precautions and Materials List

3. Test Preparation

4.0 Depletion Voltage Measurement and Laser Scan

4.1 Stage Calibration

4.2 Depletion Voltage and I-V Curve Measurement

4.3 Laser Scan

5 ADC Output Sampling and "Cal-Inject" Test

6 Instructions for Creating a Test Report

 

1.0 Test Overview

1.1 The laser test station and the PROTO test software are designed to

perform the following tests for the silicon microstrip detectors

1.1.1 Laser scan test

1.1.2 Detector ADC output sampling in Data mode

1.1.3 Detector ADC output sampling in Cal-inject mode

1.1.4 Short electrical test (using 1.1.3).

 

2.0 Safety Precautions and Materials List

2.1 Safety Precautions

2.1.1 Refer to the FERMILAB Safety Guide and federal, state, and local

government regulations.

2.2 Classified.

2.3 Materials List

2.3.1 A SaSeq board

2.3.2 A Greenboard

2.4.1 An HDI to test, by itself or bonded with sensor(s).

2.3.4 A Bit 3 crate

2.3.5 An x86 PC with Bit3 Interface card

2.3.6 "Proto" software

2.3.7 Tektronix PS2521G three-output DC voltage supply

2.3.7 Model 617 multi-source programmable voltage supply.

3.0 Test Preparation

3.1.1 Bias the detector sensor with Model 617 multi-source voltage supply

appropriately for the type of the detector being tested. Consult

your shift captain regarding the specifics.

3.1.1 Connect the Hiroshi strip terminal from the HDI to either chain A or

chain B socket on the Greenboard.

3.1.2 Supply the power the HDI by turning on the Tektronix PS2521G

triple output power supply. The power must be turned on in the

following sequence to prevent latch up of HDI circuits and possible

damage to the HDI: Memory 1, Memory 2, and Memory 3.

3.1.3 Turn on the VME board power and the fan below it.

3.1.4 On the interfaced PC, start the "PROTO" program. It can be started

by double clicking its icon or by directly invoking its full path,

usually "C:\Program Files\Proto\Proto.exe".

3.1.5 In the Proto window, choose Bit3 folder from the upper binder.

Reset the Bit3 controller by clicking on "Reset Bit 3" button.

Warning: Resetting Bit3 will destroy robot calibration. Make

sure that the "Reset Status" light turns into green when done.

If it turns red, verify that the VME board is turned on.

3.1.6 In the Proto window, choose SaSEQ/Chips folder from the upper binder.

Choose ScanView from the lower binder.

3.1.7 In the SaSEQ/Chips folder, choose the appropriate HDI type under the

"SaSEQ" listbox. To change SaSEQ configuration, click on the Setup

button above the listbox.

3.1.8 Choose the chain to which the SVX chips to test are connected in the

"Chain" list box.

3.1.9 If it is required to download a different set of parameters to the

chips than the one provided by default, select the chip in the "Chip"

list box and select the parameters to change in the parameters list

box below. Then the program will prompt for a different value for

the parameter. It automatically checks for the upper and lower

bounds of each parameters. The bounds are listed in the parameters

list box.

3.1.10 Click on the "Download Chain" button to initialize the SaSEQ and to

download the chip parameters to the chips in the selected chain.

The "Error Light" should turn green after a successful download.

If it turns red, verify that the correct SaSEQ parameters and HDI

types are selected. Also verify that the connection of the HDI to

the Green board is secure.

3.1.11 The "Scan View" window should now display a live graph of the data

outputs from all chips connected to the selected SaSEQ board.

Notice the setpoint and actual reads per second in the Scan View

Setup window frame.

 

4.0 Depletion Voltage Measurement and Laser Scan

4.1 Stage Calibration

4.1.1 At the home position of the stage, the laser lens is placed well in

front of the of the edge of the sensor away from the HDI and well in

left of the left sensor edge. If parts of the stage arm is likely to

encounter obstacles on its course to the home position, then clear

the HDI and the green board and other removable items from the stage

platform, home the stage, and then set-up the removed items again.

Be sure to turn off and on the power in the proper sequence when

removing and restoring the items.

4.1.2 Place the detector on the testing pad such that the metal heat sink

pad on bottom of the HDI carrier is securely leaning against the

right and front posts on the testing pad.

4.1.3 Choose the correct ladder type and the chain in the SaSEQ/Chips

folder. Click on "Download Chain" to download SVX chip parameters

and to initialize SaSEQ.

4.1.4 Go to "Barrido Automαtica Del Laser" folder and click on the

"Teach Robot / Scan" button. "Barrido..." stands for "Automatic

Laser Sweep".

4.1.5 In the Laser Scan control window, click on the Bit3 Init button

to initialize the Bit 3 controller if it has not previously been

initialized.

4.1.6 Click on the "Stage" button in the "Initialization" frame to reset

and home the stage. Watch where the stage is moving. To stop the

movement anytime, push the big yellow "Stop" button. If it doesn't

work, click the big red "Emergency Abort" button. If that doesn't

work either, turn off the VME crate. As a last resort, switch the

stage power switch off.

4.1.7 Click on the "SaSEQ" button in the "Initialization" frame to

initialize SaSEQ and to download configuration parameters to each

chip in the selected chain if it was not done previously.

4.1.8 Click on the "Bias" button in the "Initialization" frame to reset

the bias voltage pods. This will also turn off the pods. In the

"Bias" frame, select the appropriate supply source for the ladder

(usually #7), click on the "On/Set" button, and enter the desired

bias voltage. The bias can be turned off by resetting the bias

voltage pods again. For stage calibration purposes, higher of the

30V or the depletion voltage plus 15V is appropriate.

4.1.9 Turn on the laser. Make sure that the "Delay" knob is set to

50ns.

4.1.10 Click on "Find Best Area" button. The system will move the laser

away from the sensor and take time average of each channels.

It will then find two cleanest sensor regions for use in stage

calibration.

4.1.11 Click on the "Calibrate" button. The laser will move to four

locations in the sensor to determine 1) the spacing between each

strip, 2) the coordinate of the left corner on the sensor away

from the HDI, and 3) tilt of the sensor relative to the stage.

4.1.12 Observe the shape of the laser beam profile in the "Scan"

graph window. If the signal appears to be too weak or too

strong (saturating), press the "STOP" button when the laser

beam is still visible in the Scan graph and adjust the

luminosity of the laser. Then re-do the calibration. Do not

use the "Delay" knob to adjust the luminosity. Leave it at 50ns.

 

4.2 Depletion Voltage and I-V Curve Measurement

4.2.1 Determine the voltage scan range for the depletion voltage and

I-V curve measurement as follows using previously measured

parameters in the traveler.

Previously measured parameters:

VdTS - Depletion voltage previously measured with C-V curve.

VmMax - Maximum allowable P-side bias for double-sided

detectors.

4.2.1.1 3-Chip and H-Disk:

Positive source scan range: 0V to greater of 50V or

VdTS+20V but less than 90V.

4.2.1.2 6-Chip, 9-Chip, and F-Disk: These needs to be split-biased as

follows:

Negative source: Fix it to the higher of (in absolute value)

VmMax or (VdTS+20)/2.

Positive source scan range:: 0V to VdTS+20 minus

(the absolute value of) the Negative source voltage

but less than 90V.

4.2.2 Click on the "Parametros del Barrido" button in the "Barrido

Automαtica Del Laser" folder.

4.2.3 Type in the range for the voltage scan as determined from above

in the "Voltage Scan Range" fields.

4.2.3 Click on "OK" button to close the window.

4.2.4 In the "Laser Scan" window, click on the desired region on the

detector to measure the I-V and depletion voltage in the

"Depletion Voltage Scan" box, and type in the strip number

at which to center the laser spot. By default, the software

integrates output from 60 strips around the center strip.

It is best to choose a region in the detector free of the

noisy channels. The detector region is determined as follows:

3-Chip 6-Chip 9-Chip

Active Front Active Chips 0-2 Active, Chips 0-5

Active Back N/A N/A Active, Chips 6-8

Passive Front Passive Chips 3-5 Passive, Chips 0-5

Passive Back N/A N/A Passive, Chips 6-8

 

H-Disk F-Disk (8-Chip) F-Disk (6-Chip)

Active Front Active Chips 0-7 Chips 0-5

Active Back N/A N/A N/A

Passive Front Passive

Passive Back N/A N/A N/A

4.2.5 Make sure that the bias supply source in the Laser Scan

window is set to the positive bias source (usu. #7).

It is this source that the voltage will be stepped.

(It is not recommended to do the scan with the negative

voltage source without supervision of the shift-captain,

as the P-side of the detector can draw huge amount of

currents through the microdischarge effect)

Press "Scan Vd" to start the test.

4.2.6 During the test, if the system cannot reach a set-point

voltage, increase the voltage tolerance by (usu. from 3V

to 3.5V or 4V) in the Automatic Laser Scan Setup window

by clicking on the "Parametros de Barrido". Then press

OK.

4.2.7 To stop the voltage scan before it completes, or if the

scan stalls for the reasons other than tight voltage

tolerance, press the ">> STOP <<" button.

4.2.8 When the test is completed, a pop-up window appears with

the results of the scan. The results can be copied to the

clipboard and be used for further analysis. A section

below explains how to use Ron Lipton's Excel analysis

program to process the results.

4.2.9 Instructions for Reading the Voltage Scan Results

N - Number of data samples taken per voltage step.

VdSetPt- Setpoint voltage

VdAvg - Average value of the actual (monitored) voltage

VdStdev - 1 sigma spread in the monitored voltage

IdAvg - Average value of the monitored current

IdStdev - 1 sigma spread in the monitored voltage

AreaAvg - Average integrated sum of the ADC output

around the scanned strip. By default,

30 channels left and right of it are summed.

When the area reaches saturation, the charge

collection efficiency is at maximum.

AreaStdev - 1 sigma spread in the ADC integrated sum.

4.2.10 For a good detector, the current should stay below 30ľA in

the entire scanned voltage range. If the current is above

50ľA, then a broken capacitor is suspected and the device

should be subject to inspection by a debugging personnel.

 

4.3 Laser Scan

4.3.0 Find the Operating voltage for the detector by performing

the depletion voltage scan (per 4.2 and 6.2) if it is not

known already.

4.3.1 Set the sensor bias as follows (cf. 4.2.1 and 4.2.9).

3-Chip and H-Disk: Positive Bias: Operating voltage

6-Chip, 9-Chip, and F-Disk: These needs to be split-biased as

follows:

Negative source: Fix it to the higher of (in absolute value)

VmMax or (VdTS+20)/2.

Positive source scan range:: 0V to Operating Voltage minus

(the absolute value of) the Negative source voltage

but less than 90V.

4.3.2 Scan across the sensor by pressing "Scan Active" or "Scan Passive"

button. The stage steps across each strip, and takes a default of

10 events per strip.

4.3.3 During any point of the calibration or the laser scan, the stage

can be stopped by pressing the "STOP" button. Pressing the

"Emergency Abort" button will force abort of the stage movement

and turn off the bias voltage as well.

4.3.4 Observe the placement of the laser peak at the starting strip.

The peak of the laser should land on the first strip to be scanned.

4.3.5 After completion of the scan, view the results in the "Scan Results"

list box. The results can be copied to the Clipboard for further

processing (e.g. using Ron's data analysis software, explained in

a later section) by clicking on the "Copy To Clipboard" button in

the "Barrido Automαtica Del Laser" folder.

 

4.3.6 Instructions for Reading the Laser Scan Results

4.3.6.1 All values are baseline-suppressed.

4.3.6.2 Strip- Sensor strip number that begins from 0.

4.3.6.3 Chip - Index to the SVX chip to which the strip belongs to.

4.3.6.4 Channel - SVX channel to which the strip is connected to. This

number begins from 0.

4.3.6.5 Avg - Average value of the strip output when laser was shined directly

above it. By default, a scan takes 10 events per strip.

4.3.6.6 Avg Pass/Fail - "Failed" if the average output is outside a pre-set

bound. The bounds are set to sort out the dead, weak or overly

sensitive strips.

4.3.6.7 Stdev - Standard deviation of the scanned data.

4.3.6.8 Stdev Pass/Fail - "Failed" if the standard deviation of the output

from the channel is outside an upper limit. It is useful for

determining noisy channels.

4.3.6.9 N Sigma - The number of standard deviations the average output for a

strip is away from the strips in the sensor ranking from 25% to 75%

in magnitude. The top and bottom 25% of the data are thrown away to

avoid contamination of the data pool from non-representative strips.

A strip with high N Sigma had responded very differently during the

scan than the rest of the strips in the sensor. The mean and stdev

thus determined is referred to as the "natural excited response" of

the sensor.

4.3.6.10 N Sigma Pass/Fail - "Failed" if the output from the strip is

away from the natural sensor mean by greater of 4 natural sigmas

or 5 counts by default.

5.0 ADC Baseline Output Sampling and "Cal-Inject" Tests

5.1 These tests can be done without the aid of the stage movement unit

(i.e. without stage calibration). Be sure to turn the laser off

when performing these tests.

5.2 Click on the "Setup" button next to the "SaSeq" listbox to

open the "SaSEQ Parameter Setup" window.

5.3. Set the acquisition mode for the detector to either "Data" mode

for Baseline output sampling or to "Cal-Inject" mode for

"Cal-Inject" test.

5.4 Click on the "Commit Setpoint" button.

5.5 If "Cal-Inject" mode is selected, set the "CalVolt" parameter

appropriately to prevent signal saturation. Typical range for

testing n-side is &HE0 to &HFF. For p-side, &H00 to &H20.

Click on the "Commit Setpoint" to commit the setting and

observe the resulting signal.

5.6 Press the "Close" button in the "SaSEQ Parameter Setup" window.

5.7 Set the bias for the sensor (if necessary) by using the bias

board controls in the "Laser Scan" window invoked by pressing

"Teach Robot/Scan" button in the "Barrido Automatica Del Laser"

folder.

5.8 Reset the event statistics counter by clicking "Reset Baseline Stat"

button in the "Baseline Results" folder. The number of accumulated

events are shown in the "Total Reads" text box in the "SaSEQ/Chips"

folder.

5.9 Choose whether to suppress the offset or not by clicking on

"Suppress Offset" button in the "Scan View Setup" box.

Typically, the offset is chosen not to be suppressed. This

choice can be made anytime during the sampling, as the system

keeps track of both offset-suppressed and offset-included

statistics internally.

5.10 When enough events have been accumulated, go to "Baseline Results"

and click on "Update" button. The event statistics listbox should

now be filled with the statistics. Clicking "Update" does not

reset the event statistics, so you can keep clicking "Update"

to see progressively cumulative statistics.

5.11 The data can be exported to Clipboard for further analysis (i.e.

using Ron's data analysis software, cf. 6.0) by clicking on the

"Copy to Data to Clipboard" button in the "Baseline Results"

folder.

5.12 Instructions for Reading the Baseline Statistics Results

5.12.1 Strip- Sensor strip number that begins from 0.

5.12.2 Chip - Index to the SVX chip to which the strip belongs to.

5.12.3 Channel - SVX channel to which the strip is connected to. This

number begins from 0.

5.12.4 In each event read cycle, the program determines the natural mean and

standard deviation of the data set consisting of first 40 strips whose

readouts are closest to the median readout in each chip. The values

are then taken as the natural event statistics for the chip, as all

outliers would be excluded in the process.

5.12.5 Data - Value of the last read data for the strip.

5.12.6 Sigma - Sigma level on which the last read data stands. It is

obtained by dividing by the natural standard deviation of chip (to

which to strip belongs) the difference between the last read

data and the natural mean. Higher the Sigma, the more unusual the

strip among all other strips in the chip.

5.12.7 Mean - Mean of the accumulated events for the strip.

5.12.8 Stdev - Standard deviation of the accumulated events for the strip.

5.12.9 #oocL - Out of control low - Number of events falling below a

sigma level limit for the chip that the strip belongs to. The sigma

level limit is set to the higher of four or the number 5.

(i. e. limit for Value-Mean is the Lager of 4*Sigma or 5).

See 5.12.6 for an explanation of the sigma level.

5.12.10 #oocH - Out of control low - Number of events falling above a

sigma level for the chip that the strip belongs to. See 5.12.8.

5.12.11 % failure - Percent of the accumulated events the strip gave an

output above or below the sigma level limit.

(i. e. (#oocL+#oocH)/(total events) in percent)

5.12.12 Classification - The strip is classified as a "Failed" strip if

the % failure exceeds a certain limit, set to 5% by default.

 

6.0 Instructions for Creating a Test Report

6.1 Report Setup and Overview.

6.1.1 In the Windows Shell background window, open Ron Lipton's

"Laser_Analysis.XLS" excel analysis file.

6.1.2 Activate the "Main" sheet.

6.1.3 Click on "List failed channels and Save Files".

In the dialog that appears, enter the full Ladder ID, test

date (in m/d/y format), maximum negative bias voltage (if

known and applicable) and your name. Also select the ladder

type. Then click on "Done" button.

6.1.4 The following data are to be included in the full report:

a. Depletion voltage and IV curve scan

b. Laser scan results

c. Cal-inject (optional or if requested by shift captain)

d. Baseline (optional or if requested by shift captain)

6.1.5 When the last piece of data is entered into the report,

save the data on the disk by clicking on "List failed channels

and Save Files" button in the "Main" sheet. A second copy of

the report will be saved in "\\D0SERVER4\PROJECTS\SIDET\LASERTEST"

remote folder.

6.1.6 The definition of Active and Passive region of the detector

is as follows for each detector type when creating this report:

3-Chip 6-Chip 9-Chip

Active Front Active Chips 0-2 Active, Chips 0-5

Active Back N/A N/A Active, Chips 6-8

Passive Front Passive Chips 3-5 Passive, Chips 0-5

Passive Back N/A N/A Passive, Chips 6-8

 

H-Disk F-Disk (14-Chips)

Active Front Active 8-chip side

Active Back N/A N/A

Passive Front Passive 6-chip side

Passive Back N/A N/A

6.2 Creating Depletion Voltage and IV Scan Report

6.2.1 When an IV scan is done to create I-V curve and to measure

the depletion voltage scan, copy the data to the clipboard.

(per 4.2.8)

6.2.2 Paste the data on the cell "A1" of the appropriate sheet in the

analysis spreadsheet, as shown below:

Active region - Paste data to "VdActiveFront" sheet.

View the chart in the "VdActiveFrontChart" chart.

Passive region - Paste data to "VdPassiveFront" sheet.

View the chart in the "VdPassiveFrontChart" chart.

P-side IV curve - Paste data to "PSideIV" sheet.

View the chart in the "PSideIV" chart.

6.2.4 View and print the IV curve and depletion voltage scan chart by

activating the chart sheet and activating "Print" command under

"File" menu.

6.2.5 To determine Vd, draw a best-fit line fitting the

Area-Voltage curve from the lowest scanned voltage (usu. 0V).

Only include the linear region and exclude the region where

the slope bends appreciably. Then draw another best

fit line along the region where Area-Voltage curve saturates.

Intercept of the two lines gives the Vd as measured by the

laser scan. Make sure to add the negative bias voltage to

the result to get the correct Vd for double-sided sensors.

6.2.6 To determine the Operating Voltage, find a reasonably

saturated region in the Area-Voltage curve within

Vd+10 to 25V, with the absolute maximum of 90V.

Make sure to add the negative bias voltage to the result

to get the correct Operating Voltage for the double-sided

sensors. This voltage is to be used in the laser scan.

 

6.3 Creating a Laser Scan Report

6.3.1 When each scan is finished (Passive or Active) for the detector,

copy the data to the clipboard (per 4.3.5).

6.3.2 Activate either "Passive" or "Active" sheet in the analysis

spreadsheet. Select cell "A1". Then click on "Paste" command

under "Edit" menu to paste the data.

6.3.3 View the graph by clicking on "ActiveChart" for the active region

chart, or "PassiveChart" for the passive region chart.

6.3.4 Print the charts by invoking on "Print" command in "File" menu.

6.3.5 Click on "List failed channels and Save Files" in the "Main" sheet

to list the dead channels, which are the channels that gives ADC

output of less than 40 above the pedestal when shined to by the

laser.

6.3.6 Print the list by invoking on "Print" command in "File" menu.

6.3.7 Instructions for Reading the Scan Chart

6.3.5.1 Each graph shows the average and the standard deviation of the

pedestal-suppressed response of the each channel when the laser

was shined on top of it. The scan program takes a default of 10

measurements per strip.

6.3.5.2 A dead channel does not respond to the laser, so its pedestal-

suppressed response will be close to the zero

6.3.5.3 Also note the channels that responds to the laser weaker or stronger

than their peers. Those channels that doesn't fit into distribution

tend to lie in a contiguous region, and tend to have high standard

deviation in the response as well. They represent defects within the

sensor.

 

6.4 Creating a Cal-inject and Baseline Readout Report

6.4.1 Copy the data to the clipboard after a cal-inject or a baseline

readout is done (per 5.0).

6.4.2 Paste the cal-inject data to the cell "A1" of "Cal-Inject" sheet, and

baseline readout data to the cell "A1" of the "Baseline" sheet.

6.4.3 View the graphical output in "Cal-Inject Chart" and "BaselineChart"

chart sheets. Print them.

 

6.5 Archiving the Report

6.5.1 Insert the printed Vd/IV charts, laser scan charts, and dead-channel

list in the detector traveler.

6.5.2 Fill out the laser scan test report in the detector traveler.

6.5.3 Fill out the laser scan test report in the electronic database

in "http://martureo.fnal.gov/d0smtproduction/database/database.asp"

page.

6.5.4 At the end of the shift, review the finished scan reports with the

shift captain for detector disposition.

 

7.0 Software Parameter Configuration

7.0.1 All parameters for configuring the software is set in the "Automatic

Laser Scan Setup" window. The window is invoked by pressing

"Parαmetros Del Barrido" button in the "Barrido Automαtica Del Laser"

folder.

7.1 Short Electrical Test Parameters

7.1.1 The parameters for short electrical test are in "SPC for Baseline

Output Noise" frame window.

7.1.2 The Negative and Positive delta limit sets a cut (in multiples of

natural chip sigma) for the output from each channel. The minimum value

for such cut in either positive or negative direction from the natural

mean is set the in the Minimum Positive or Negative Delta.

7.1.3 "Failure Tolerance" - The maximum percent of the time a strip can

give the output outside the cuts without being labeled as failed.

7.1.4 "Minimum Data Set" - Baseline Results will not be computed unless a

minimum of this number of events have been processed.

7.2 Laser Scan Test Parameters

7.2.1 The parameters for the laser scan test are in the "Configuration at

Scan" frame window.

7.2.3 Channel Output Specs - Acceptable ranges for the average and the

standard deviation of the collected samples during the scan.

7.2.4 SPC Specs - Maximum allowed deviations for an averaged strip output

to be from the natural mean of the sensor. This limit is larger of

a constant or a multiple of the natural sensor standard deviation.

7.2.1 Samples Per Scan - Number of events to take when the laser is shined

on the strip. Laser will move to the next strip after that.

7.2.2 Max Scan Retries - Maximum number of attempts to retake the samples

when either there is an I/O error or when the scan result for the

channel is outside the specification. All good reads prior to the

retry for the strip is discarded, and the events are taken anew.

7.3 Other Notes

7.3.1 The program powers and enables both chains at the same time by default.

To enable only one chain, set the IGNORE register in SaSEQ parameter

setup dialog box as follows and press "Commit" button.

Chain A only: &H20

Chain B only: &H10

Both chains (default): &H0

7.3.2 The program sends the following sequence of commands to the SaSEQ

TRIGGER register to acquire the signal:

&H00: ;Set acquire mode with 132ns crossing time

&H02 &H02 &H02 ;Sample with 132ns crossing time

(Performs a sequential read)

&HA8 ;Initialize

CSR to &HA6 ;Reset SaSEQ board

 

==============================================================================

Revision History

==============================================================================

Ver | Date | Name | Reason

----+--------------+--------------------+-------------------------------------

001 Oct 04, 1999 John Rha New Document

002 Nov 03, 1999 John Rha Section 4.1- Document update for the

scan software release 1.6.

1. Added instructions for improved

stage calibration algorithm.

2. Added sensor bias control

instructions.

3. Added data export facility to

Ron Lipton's laser test report

generator Excel spreadsheet.

003 Nov 12, 1999 John Rha Section 4.1-

1. Required to home the stage

before removing the detector

from the laser test station.

2. Required to adjust laser

luminosity to avoid weak signal

or signal saturation.

004 Jan 12, 1999 John Rha Section 4 -

1. Added procedures for IV-curve

generation and depletion

voltage scan.

2. Added detector biasing

instructions.

Section 5 -

1. Added cal-inject instruction.

Section 6 -

1. Updated test report generation

instruction to accomodate an

enhanced version of the Ron's

report generator spreadsheet.

==============================================================================